Sign in →

Test Code VBG BLD GAS,VENOUS

Additional Codes

LAB3032

Performing Laboratory

Porter Hospital Laboratory

Specimen Requirements

Specimen must arrive within 30 minutes of draw.
Specimen Type: Whole blood-Clotted blood is not acceptable.
Container/Tube: Lithium heparin plasma seperator tube (PST)
Specimen Volume: 2 mL

Collection Instructions:
1. Draw VENOUS blood anaerobically.
2. Expel air bubbles, remove needle from syringe and cap-Air bubble is not acceptable.
3. Ideally, the sample should be tested within 15 minutes of collection. If this cannot be achieved, place on ice for up to 2 hours.

Test Schedule / Analytical Time / Test Priority / Turn Around Time

Daily / Immediately / Available STAT

Methodology

Instrument: RapidPoint 

pH               Potentiometric method using standard ionselective electrode (ISE) technology
pCO2          Modified potentiometric method based on the principle of the Severinghaus electrode
pO2           Amperometric measurement based on the principles of the Clark electrode

Test Classification and CPT Coding

82803

Normal Reference Values

VENOUS Blood Gas Reference Intervals - ALL AGES

Analyte

 

Units

 

pH

7.31-7.41**

 

 

pO2

None

 

 

pCO2

41-51**

mmHg

 

HCO3

23-28**

mmol/L

 

O2 Sat

None

 

 

Lactate

0.36-1.39***

mmol/L

 

BE

(-2) to (+3)

mmol/L

 

** iSTAT

***RapidPoint

 

Critical Values

None Established